Meter Proving Repeatability Calculator

API MPMS Chapter 4 & Chapter 12 Proving Analysis

Meter Proving Repeatability Calculator (API MPMS Ch. 4)
Enter proving run data to calculate meter factor, evaluate repeatability against the 0.05% criterion, detect statistical outliers, and determine proving validity. Supports both volume-based and pulse-based proving data entry.

Proving Run Data

Enter 3-10 consecutive proving run values (meter factors or K-factors).

%

Proving Configuration

Operating Conditions (Optional)

°F
psig

Understanding Meter Proving

Meter Factor
MF = Prover Volume / Meter Volume. Values near 1.0000 indicate good agreement. Applied as correction to meter readings.
Repeatability:
Repeatability = (Max MF - Min MF) / Mean MF × 100%. Must be ≤0.05% per API MPMS Ch. 4 for valid proving.

Formula

Repeatability = (MF_max − MF_min) / MF_avg × 100
MF = Meter Factor (Prover Vol / Meter Vol)
Criterion = Repeatability ≤ 0.05%
Runs = Minimum 5 consecutive (API MPMS Ch. 4)

Standards & References

  • API MPMS Chapter 4
    Proving Systems
  • API MPMS Chapter 12
    Calculation of Petroleum Quantities
  • API MPMS Chapter 13
    Statistical Aspects of Measuring and Sampling
  • AGA Report No. 7
    Measurement of Gas by Turbine Meters

Engineering Notes

  • 0.05% rule: Range of meter factors must not exceed 0.05% of the mean for valid proving
  • Minimum runs: API requires minimum 5 runs for compact provers, 3 for pipe provers
  • Outliers: Grubbs test used to detect statistical outliers at 95% confidence
  • Trending: Track meter factor over time to detect meter wear or calibration drift