Pipeline Operations — Measurement

Meter Proving Fundamentals

Meter proving is the process of determining a meter factor by comparing the meter's indicated volume to a known reference volume. It is the cornerstone of custody transfer accuracy, providing the traceable link between field measurement and standard conditions. Understanding proving procedures, repeatability criteria, and statistical analysis is essential for measurement technicians and engineers.

Meter Factor

MF = V_prover / V_meter

Ratio of prover reference volume to meter indicated volume.

Repeatability Criterion

≤ 0.05%

API MPMS Ch. 4. Range/Mean of consecutive runs.

Key Standards

API MPMS Ch. 4 · Ch. 12

Proving systems and calculation of petroleum quantities.

Use this guide when you need to:

  • Calculate meter factor from proving data.
  • Evaluate proving repeatability against API criteria.
  • Detect and handle statistical outliers.
  • Establish proving frequency and procedures.

1. Purpose of Meter Proving

Every flow meter has inherent inaccuracies from manufacturing tolerances, installation effects, and operating condition variations. Meter proving quantifies these inaccuracies by comparing the meter's output to a reference standard of known accuracy. The resulting meter factor is then applied to all subsequent meter readings to correct for the systematic error.

Proving is required at custody transfer points because the financial value of the measured fluid justifies the cost of maintaining traceable measurement accuracy. A 0.1% error on a station flowing 100 MMSCF/d at $3.50/MMBTU represents approximately $130,000 per year of undetected measurement bias.

Without Proving

Unquantified Error

Meter reads within manufacturer spec (typically 0.5-2.0%) but actual bias is unknown.

With Proving

Corrected Measurement

Meter factor corrects bias to within proving uncertainty (typically 0.02-0.05%).

Financial Impact

$130k+ per 0.1%

On a 100 MMSCF/d station at $3.50/MMBTU, every 0.1% matters.

Traceability chain: The prover's reference volume is calibrated against a primary standard (water draw, gravimetric), which is traceable to NIST. This unbroken chain of calibration provides the legal basis for custody transfer measurement accuracy.

2. Meter Factor

Meter Factor Definition: MF = V_prover / V_meter Where: MF = Meter Factor (dimensionless) V_prover = Volume indicated by prover (reference) V_meter = Volume indicated by meter under test Applied as correction: V_corrected = V_meter_indicated × MF Example: Prover indicates: 100.12 gallons (calibrated reference) Meter indicates: 100.00 gallons (under test) MF = 100.12 / 100.00 = 1.0012 All subsequent meter volumes are multiplied by 1.0012 to correct for the meter's slight under-reading. For K-factor (pulse) meters: K = Pulses / V_prover MF = K_base / K_proving (ratio to baseline K-factor)

Meter Factor Interpretation

MF ValueMeaningTypical Range
MF = 1.0000Meter reads exactly correctIdeal (rare)
MF > 1.0000Meter under-reads (correction adds volume)Common for turbine meters
MF < 1.0000Meter over-reads (correction reduces volume)Common for new meters
MF > 1.02 or < 0.98Significant deviation, investigateMay indicate problem

3. Prover Types

Prover TypeVolumeMin RunsTypical RepeatabilityApplication
Pipe Prover10-300 bbl30.01-0.02%Large liquid stations, permanent installation
Compact Prover0.01-1 bbl50.02-0.04%Smaller stations, portable proving
Master MeterN/A50.02-0.05%Gas metering, transfer standard
Tank Prover100-5000 gal30.01-0.02%Primary calibration, water draw
Compact vs. pipe provers: Compact provers use a small piston and many passes to accumulate sufficient volume. Because each pass has higher relative uncertainty, API requires minimum 5 runs (vs. 3 for pipe provers). The advantage is portability and smaller footprint.

4. Repeatability Criterion

API MPMS Chapter 4 Repeatability: Repeatability (%) = (MF_max - MF_min) / MF_avg × 100 Criterion: Repeatability must not exceed 0.05% Example: Run 1: MF = 1.0012 Run 2: MF = 1.0015 Run 3: MF = 1.0010 Run 4: MF = 1.0014 Run 5: MF = 1.0011 MF_avg = 1.00124 MF_max = 1.0015, MF_min = 1.0010 Range = 0.0005 Repeatability = 0.0005 / 1.00124 × 100 = 0.0499% Result: PASS (0.0499% < 0.05%) MF to apply: 1.00124 (mean of all 5 runs)

If repeatability exceeds 0.05%, the proving is invalid. The technician must identify and correct the cause before re-proving. Common causes include air or gas entrainment, temperature changes during proving, prover valve leaks, meter malfunction, or insufficient flow stabilization.

5. Outlier Detection

Statistical outlier detection identifies individual proving runs that are significantly different from the others. The Grubbs test is the most commonly used method in custody transfer metering. It compares each value's deviation from the mean to the standard deviation, testing against critical values for the sample size.

Grubbs Test: G = |x_suspect - x_mean| / s Where: G = Grubbs test statistic x_suspect = the value being tested x_mean = sample mean s = sample standard deviation Decision rule: If G > G_critical (from table for N and alpha=0.05), the value is an outlier and may be excluded. Critical values (alpha=0.05, two-sided): N=3: G=1.155 N=5: G=1.715 N=7: G=2.020 N=10: G=2.290 Important: - Only remove outliers with identifiable physical cause - Re-prove after removing outlier run - Never remove more than one outlier from a set - Document reason for exclusion

6. Proving Procedure

  • Stabilize conditions: Allow flow rate, temperature, and pressure to stabilize before starting. Temperature should not vary more than 1 deg F during a run.
  • Verify prover: Check prover calibration certificate, valve seal integrity, and detector switch operation before proving.
  • Run proving: Perform minimum required number of consecutive runs without interruption. Record meter pulses and prover volumes for each run.
  • Check repeatability: Calculate repeatability after each run. If criterion is met with minimum required runs, proving is valid.
  • Calculate meter factor: Average all accepted run meter factors. Apply temperature, pressure, and compressibility corrections as applicable.
  • Document: Record all data, conditions, meter factor, and any anomalies. Retain records per contract and regulatory requirements.

7. Volume Corrections

Volume Correction Factors: V_standard = V_indicated × Ctl × Cpl × Ctm × Cpm × MF Where: Ctl = Temperature correction for liquid (prover) Cpl = Pressure correction for liquid (prover) Ctm = Temperature correction for meter Cpm = Pressure correction for meter MF = Meter factor from proving For gas metering: V_standard = V_actual × (P/P_base) × (T_base/T) × (1/Z) × MF Where Z = compressibility factor (AGA 8 / SGERG-88)

8. Proving Frequency

ApplicationTypical FrequencyTrigger Events
Custody transfer (liquid)Monthly or per batchProduct change, meter maintenance, MF shift >0.0025
Custody transfer (gas)Monthly to quarterlyCalibration, flow profile change, maintenance
Allocation meteringQuarterly to annuallySignificant flow or composition change
Check meteringAnnuallyWhen discrepancy with primary meter exceeds tolerance

9. Troubleshooting Poor Repeatability

SymptomPossible CauseSolution
Repeatability > 0.05%Air/gas entrainmentVent prover, check for leaks upstream
Systematic drift in MFTemperature change during provingStabilize conditions, insulate lines
Random scatterProver valve leakingTest prover valve leak rate
One run differentTransient flow disturbanceExclude outlier (with documentation), re-run
All MFs shifted from previousMeter wear or contaminationInspect meter internals, clean, recalibrate

10. Industry Standards

StandardTitleRelevance
API MPMS Ch. 4Proving SystemsProver types, procedures, repeatability
API MPMS Ch. 5MeteringMeter types and specifications
API MPMS Ch. 12Calculation of Petroleum QuantitiesVolume correction factors
API MPMS Ch. 13Statistical Aspects of Measuring and SamplingOutlier detection, data analysis
AGA Report No. 7Measurement of Gas by Turbine MetersTurbine meter proving requirements
AGA Report No. 9Measurement of Gas by Ultrasonic MetersUSM proving and diagnostics
API MPMS Ch. 21Flow Measurement Using Electronic Metering SystemsFlow computer requirements