1. GC Operating Principles
A gas chromatograph separates a gas mixture into its individual components by exploiting differences in how each component interacts with a stationary phase inside a chromatographic column. The separated components are then detected and quantified by one or more detectors. For natural gas analysis, the GC separates methane, ethane, propane, butanes, pentanes, hexanes+, nitrogen, and carbon dioxide, and optionally hydrogen sulfide, oxygen, and other trace components.
Basic GC Process
Carrier Gas
The carrier gas transports the sample through the column without reacting with it. Helium is the traditional carrier gas for natural gas GCs because it has high thermal conductivity (important for TCD sensitivity) and is inert. Hydrogen is increasingly used as an alternative due to helium shortages and cost; hydrogen provides faster analysis times but requires additional safety considerations for explosive atmosphere classification.
Helium Carrier
Standard Choice
High TCD sensitivity, inert, safe. Increasingly expensive. Flow rate typically 20-30 mL/min.
Hydrogen Carrier
Cost-Effective Alternative
Faster separations, lower cost. Requires Class I Div 1/2 considerations for safety. Used with FID detectors.
Nitrogen Carrier
Rarely Used
Poor TCD sensitivity for light hydrocarbons. Only used when N2 is not a required analyte.
2. Detectors (TCD & FID)
Natural gas GCs primarily use two detector types: the Thermal Conductivity Detector (TCD) for universal detection of all components, and the Flame Ionization Detector (FID) for enhanced sensitivity to hydrocarbons.
Thermal Conductivity Detector (TCD)
The TCD measures the difference in thermal conductivity between the pure carrier gas and the carrier gas plus eluted component. Because every gas has a characteristic thermal conductivity, the TCD can detect all components in natural gas including inert gases (N2, CO2). When using helium carrier, which has very high thermal conductivity, even small concentrations of other gases produce measurable signals.
Flame Ionization Detector (FID)
The FID burns the eluted component in a hydrogen/air flame, producing ions proportional to the number of carbon atoms in the molecule. The FID is roughly 1000 times more sensitive than the TCD for hydrocarbons, but it does not respond to inert gases (N2, CO2, H2O, noble gases) because they do not produce ions when burned.
Detector Configuration
Most custody-grade natural gas GCs use a TCD for the main analysis (all components) and may add an FID for improved sensitivity to trace hydrocarbons (C6+, C7+). Some modern process GCs use dual TCD configurations with different column packings to handle both light ends (C1, C2, CO2, N2) and heavy ends (C3+) in parallel channels.
3. Columns & Separation
The chromatographic column is the heart of the GC. It separates the gas mixture based on differences in how strongly each component interacts with the stationary phase. Natural gas GCs typically use packed columns rather than capillary columns because packed columns handle the larger sample volumes needed for process analysis and are more robust in field conditions.
Common Column Configurations
| Column Type | Stationary Phase | Components Separated | Typical Dimensions |
|---|---|---|---|
| Molecular Sieve 5A | Zeolite adsorbent | O2, N2, CH4 (permanent gases) | 6-12 ft × 1/8 in |
| Silica Gel / Porapak Q | Porous polymer | CO2, C2, C3 (acid gas + light HC) | 6-12 ft × 1/8 in |
| OV-101 / DC-200 | Methylsilicone on Chromosorb | C4, C5, C6+ (heavier HC) | 6-20 ft × 1/8 in |
| Alumina PLOT | Al2O3 on capillary | C1-C6+ (single column approach) | 30-50 m × 0.53 mm |
Column Switching (Valve-Based)
Natural gas GCs use multi-column configurations with valve switching to achieve the required separation of all components. A typical GPA 2261 analysis uses two or three columns in series with backflush and heart-cut valves to route components through the optimal column for separation. Column switching is controlled by precise timing based on the known elution order of the calibration gas components.
4. Calibration Gas Standards
Calibration gas is a certified reference mixture with precisely known concentrations of each component. The accuracy of every GC analysis is directly traceable to the calibration gas. For custody transfer applications, calibration gas must be a primary reference standard or traceable to a primary standard from an accredited laboratory.
Calibration Gas Requirements
Calibration Gas Handling
- Storage temperature: Store calibration gas cylinders in a temperature-controlled environment (60-100 °F). Extreme temperatures can cause component condensation (especially C5+ and C6+) that changes the gas-phase composition.
- Cylinder pressure: Do not use calibration gas below 200 psig cylinder pressure. As pressure drops, heavier components preferentially remain in the liquid phase, enriching the gas-phase composition with lighter components.
- Regulators: Use dedicated regulators for calibration gas to prevent cross-contamination. Use stainless steel diaphragm regulators, not brass.
- Line purging: Purge calibration gas supply lines for at least 5 minutes before calibration to ensure the sample reaching the GC is representative of the cylinder contents.
5. Calibration Procedure
Calibration establishes the relationship between detector response (peak area) and component concentration. For a properly operating GC, this relationship should be linear over the expected concentration range. GPA 2261 specifies a minimum calibration procedure that includes verification of component identification, response factor calculation, and repeatability checks.
Step-by-Step Calibration
Calibration Frequency
| Application | Calibration Interval | Verification Interval |
|---|---|---|
| Custody transfer (AGA/API) | Monthly or per contract | Daily or per analysis cycle |
| Process monitoring | Quarterly | Weekly |
| Environmental monitoring | Per regulatory requirement | Per regulatory requirement |
6. Response Factor Calculation
The response factor (RF) quantifies the detector's sensitivity to each component. It converts raw peak area (or peak height) to concentration. Response factors must be stable over time; significant changes indicate detector problems, column degradation, or system leaks.
Response Factor Stability
Response factors should be monitored over time by maintaining a control chart. A stable GC will show RF values that vary by less than 1-2% relative from calibration to calibration. Sudden changes in RF indicate problems that must be investigated before the GC is used for custody transfer analysis.
| RF Change | Possible Cause | Action |
|---|---|---|
| All RFs decreased uniformly | Sample loop leak, low sample pressure | Leak test sample system, verify loop pressure |
| Heavy component RFs decreased | Column contamination, cold spot in sample line | Bake out column, heat trace sample lines |
| Single component RF changed | Co-elution, peak integration error | Check resolution, adjust integration parameters |
| All RFs increased uniformly | Detector sensitivity increase, carrier flow decrease | Verify carrier gas flow, check detector settings |
7. Linearity Verification
Linearity verification confirms that the detector response is proportional to concentration over the entire measurement range. A linear detector allows accurate analysis even when the sample composition differs significantly from the calibration gas. GPA 2261 requires linearity verification as part of GC qualification.
8. Repeatability Requirements
Repeatability is the variation in results when the same sample is analyzed multiple times under the same conditions. It is the most fundamental measure of GC precision and must meet specified limits for the results to be considered valid for custody transfer.
Reproducibility vs. Repeatability
Repeatability measures variation within a single GC under identical conditions. Reproducibility measures variation between different GCs, different operators, or different laboratories analyzing the same gas. Reproducibility limits are approximately 2 to 3 times the repeatability limits. When two custody transfer GCs at the same location produce different results, the reproducibility limits determine whether the difference is within acceptable measurement uncertainty.
Heating Value Impact
The primary purpose of GC analysis at a custody transfer point is to determine the heating value (Btu/scf) of the gas for billing purposes. A typical natural gas composition change of 0.1 mol% in ethane or propane changes the heating value by approximately 1-3 Btu/scf. At a flow rate of 100 MMscf/d and $3.00/MMBtu, a 3 Btu/scf measurement error represents approximately $900/day or $328,000/year in billing error.
9. Common Troubleshooting
Field GCs operate in harsh environments and require regular maintenance and troubleshooting. The most common problems fall into categories of sample system issues, column degradation, detector problems, and electronic/software errors.
Diagnostic Table
| Symptom | Probable Cause | Corrective Action |
|---|---|---|
| All peaks smaller than expected | Sample loop leak, low sample pressure, carrier flow too high | Leak test with soap solution, verify sample loop pressure, check carrier flow |
| Heavy peaks missing or reduced | Cold spot in sample line, column contamination | Heat trace sample lines, bake out or replace column |
| Peaks tailing or broad | Column degradation, dead volume in fittings | Replace column, check and tighten all fittings |
| Baseline drift | Detector contamination, carrier gas impurity, column bleed | Clean detector, replace carrier gas trap, condition column |
| Retention times shifted | Carrier flow change, oven temperature change, column aging | Verify carrier flow and oven temp, recalibrate retention times |
| Ghost peaks / extra peaks | Sample carryover, contaminated sample lines, outgassing | Purge sample system, run blank analysis, check for contamination |
| Poor repeatability | Valve timing issue, sample pressure variation, electrical noise | Check valve actuation timing, regulate sample pressure, check grounding |
| Methane/ethane not resolved | Molecular sieve column saturated or contaminated with CO2 | Bake out molecular sieve at 300 °C, replace if needed |
Preventive Maintenance
- Sample system: Replace coalescing filters monthly. Check for liquid accumulation in sample probes and lines daily. Verify heat trace temperature on sample lines.
- Carrier gas: Replace moisture and hydrocarbon traps per manufacturer schedule (typically annually or when indicator changes color).
- Detector: Clean TCD filaments annually. Replace FID jet and igniter as needed. Verify detector temperature.
- Valves: Check valve switching timing and actuation. Replace valve seals per manufacturer schedule (typically 1-2 years).
- Column: Monitor retention times and resolution. Condition columns periodically by temperature programming to remove accumulated heavy contaminants.
10. Standards & References
| Standard | Title | Relevance |
|---|---|---|
| GPA 2261 | Analysis of Natural Gas and Similar Mixtures by GC | Primary GC analysis standard for natural gas |
| GPA 2166 | Obtaining Natural Gas Samples for Analysis by GC | Proper sampling procedures upstream of GC |
| GPA 2172 | Calculation of Gross Heating Value, Relative Density, etc. | Converting GC composition to heating value |
| ISO 6974 | Natural Gas — Determination of Composition by GC | International GC analysis standard |
| ISO 6976 | Natural Gas — Calculation of Calorific Values | International heating value calculation |
| AGA Report No. 5 | Natural Gas Energy Measurement | Energy determination from composition |
| AGA Report No. 8 | Compressibility Factors (DETAIL equation) | Z-factor from GC composition for flow calculation |
| API MPMS Ch. 14.1 | Collecting and Handling of Natural Gas Samples | Sampling requirements for custody transfer GC |
| ASTM D1945 | Standard Test Method for Analysis of Natural Gas by GC | Alternative GC analysis standard |